• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: THORATEC CORPORATION HEARTMATE II SYSTEM CONTROLLER; VENTRICULAR (ASSISST) BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

THORATEC CORPORATION HEARTMATE II SYSTEM CONTROLLER; VENTRICULAR (ASSISST) BYPASS Back to Search Results
Model Number 106762
Medical Device Problem Code Display or Visual Feedback Problem (1184)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/18/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Event or Problem Description
It was reported that the patient could not see the green arrows on their controller anymore.The controller was exchanged.No additional information was provided.
 
Additional Manufacturer Narrative
Manufacturer's investigation conclusion: the reported event of a dim pump running led was confirmed via visual analysis.The heartmate ii system controller was returned for analysis and a log file was downloaded for review.A review of the submitted log files showed 240 events spanning approximately 48 days ((b)(6) 2021 per timestamp).There were no notable events active in the log file that would indicate an issue with the system controller.The driveline was disconnected on (b)(6) 2021 at 23:51:31 for a system controller exchange.The system controller was connected to a mock loop and was able to support the system for an extended period of time without issue.The power cables of the system controller were manipulated by hand, while connected to the system, without any alarms activating.Upon connection of the power cables to the power module, the dim pump running led was observed, and the reported event was confirmed.The dim led did not affect pump operation.A root cause for the reported event was unable to be conclusively determined through this analysis.Incidental findings: damaged connector nut.The device history records were reviewed and the records revealed the heartmate ii system controller was manufactured in accordance with manufacturing and quality assurance specifications.Heartmate ii instructions for use section 8 ¿ ¿equipment storage and care¿ and heartmate ii patient handbook section 6 ¿ ¿caring for equipment¿ explain how to properly care for the equipment.Additionally, heartmate ii patient handbook section 10 entitled ¿safety checklists¿, instructs users to regularly inspect their accessories for damage, and to replace any equipment that appears damaged or worn.Heartmate ii instructions for use section 3 ¿ ¿powering the system¿ and heartmate ii patient handbook section 3 ¿ ¿powering the system¿ addresses the user to not use expired batteries and advises the user to properly dispose of them.Heartmate ii patient handbook and heartmate ii instructions for use (ifu) caution users to call their hospital contacts if they think, for any reason, any portion of their equipment is not functioning as usual, is broken, or they are uncomfortable with the operation of the equipment.No further information was provided.The manufacturer is closing the file on this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEARTMATE II SYSTEM CONTROLLER
Common Device Name
VENTRICULAR (ASSISST) BYPASS
Manufacturer (Section D)
THORATEC CORPORATION
6035 stoneridge drive
pleasanton CA 94588
MDR Report Key11995833
Report Number2916596-2021-02967
Device Sequence Number922271
Product Code DSQ
UDI-Device Identifier00813024013013
UDI-Public00813024013013
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
P060040
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date (Section B) 07/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/06/2019
Device Model Number106762
Device Catalogue Number10005080
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/11/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 05/25/2021
Supplement Date Received by Manufacturer07/22/2021
Initial Report FDA Received Date06/14/2021
Supplement Report FDA Received Date07/30/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age73 YR
Patient Weight62
-
-