| Model Number |
MS1-4570S |
| Medical Device Problem Code |
Mechanical Problem (1384)
|
| Health Effect - Clinical Code |
Failure of Implant (1924)
|
| Date of Event |
02/23/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device evaluation: the returned rod was observed to be partially distracted with score marks on the distraction rod.X-ray images of internal components showed no obvious damage and revealed the rod was partially distracted.The rod was functionally tested and could not distract and retract with the electronic remote controller (erc) and the manual distractor.Because the rod was unable to retract or distract, force testing and the stroke measurement test could not be performed.Sectioning of the rod revealed the distraction rod was contacting the internal surface of the housing.The two components could not be separated without applying high force.After applying high force, the two components they were able to be separated.The potential root cause of the reported event was due to patient activity causing the distraction rod to become bent causing a jam.Device records review: review of the device history record for the rod confirmed that it met all of the required quality inspections prior to shipment.
|
| |
|
Event or Problem Description
|
|
Information was received that a revision surgery was performed due to the surgeon believing that the rods were not deploying due to extended time between distractions.No patient harm was reported.
|
| |
|