• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: APPLIED MEDICAL RESOURCES CFS12, 11X100 KII SLEEVE ADVFIX 12/BX; LAPAROSCOPE, GENERAL & PLASTIC SURGERY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

APPLIED MEDICAL RESOURCES CFS12, 11X100 KII SLEEVE ADVFIX 12/BX; LAPAROSCOPE, GENERAL & PLASTIC SURGERY Back to Search Results
Model Number CFS12
Medical Device Problem Code Device Contaminated During Manufacture or Shipping (2969)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/23/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The event unit returned to applied medical for evaluation.A follow-up report will be provided upon completion of the investigation.
 
Event or Problem Description
Name of procedure being performed: na (incoming inspection).Detailed description of event: [name] incoming inspection po# (b)(4).This is a complaint from [name].The lot mentioned above was found to have defective units per our incoming inspection.Delivery date: april 21, 2021.We found defects as the result of incoming inspection of the products.Please refer to the attachment material for the quantity which returns the products.We found the defects that already informed applied medical by cer.We just inform you the defect and return the products for following defects.You do not need investigation.Hairlike object in pouch - cfs12 lot# 1407319 (1 ea).Patient status: patient not involved.Type of intervention: na.
 
Event or Problem Description
Name of procedure being performed na (incoming inspection).Detailed description of event: [name] incoming inspection po# (b)(4).This is a complaint from [name].The lot mentioned above was found to have defective units per our incoming inspection.Delivery date: april 21, 2021.Please refer to the attachment file.We found defects as the result of incoming inspection of the products.Please refer to the attachment material for the quantity which returns the products.We found the defects that already informed applied medical by cer.We just inform you the defect and return the products for following defects.You do not need investigation.Hairlike object in pouch - cfs12 lot# 1407319 (1 ea).Patient status: patient not involved.Type of intervention na.
 
Additional Manufacturer Narrative
The event unit was returned to applied medical for evaluation.Visual inspection confirmed the presence of hair inside the sterile unopened pouch.Based on the condition of the returned unit and the description of the event, the hair originated from an operator during the manufacturing process.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CFS12, 11X100 KII SLEEVE ADVFIX 12/BX
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
APPLIED MEDICAL RESOURCES
22872 avenida empresa
rancho santa margarita CA 92688
MDR Report Key11996132
Report Number2027111-2021-00524
Device Sequence Number14503031
Product Code GCJ
UDI-Device Identifier00607915119850
UDI-Public(01)00607915119850(17)240224(30)01(10)1407319
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K083638
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source distributor,foreign
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Expiration Date02/24/2024
Device Model NumberCFS12
Device Catalogue Number101245401
Device Lot Number1407319
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/09/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 05/18/2021
Supplement Date Received by Manufacturer05/18/2021
Initial Report FDA Received Date06/14/2021
Supplement Report FDA Received Date08/19/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-