| Model Number |
5530-G-511 |
| Medical Device Problem Code |
Unstable (1667)
|
| Health Effect - Clinical Code |
Ambulation or Postural Difficulties (2544)
|
| Date of Event |
05/12/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.There have been no other similar events for the lot referenced.An evaluation of the device cannot be performed as the device was not returned to the manufacturer.Should additional information become available it will be reported in a supplemental report upon completion of the investigation.
|
| |
|
Event or Problem Description
|
|
Revision of insert due to instability.
|
| |
|
Event or Problem Description
|
|
Revision of insert due to instability.
|
| |
|
Additional Manufacturer Narrative
|
|
Reported event: an event regarding instability involving a triathlon insert was reported.The event was not confirmed.Method & results: product evaluation and results: visual inspection of the returned device noted the following: examination by engineer indicated common damage modes of uhmwpe present on the returned device.Explantation damage was also observed on the insert.Material analysis of the returned device noted the following: ma - examination by engineer indicated common damage modes of uhmwpe present on the returned device.Explantation damage was also observed on the insert.Based on the given information, no identifiable materials or manufacturing discrepancies were observed on the surfaces examined.Clinician review: no medical records were received for review with a clinical consultant.Product history review: review of the device history records indicate devices were manufactured and accepted into final stock with no relevant reported discrepancies.Complaint history review: there have been no other similar events for the lot referenced.Conclusions: the exact cause of the event could not be determined because insufficient information was provided.Further information such as pathology reports, pre- and post-operative x-rays and the primary operative report as well as patient history and follow-up notes are needed to complete the investigation for determining root cause.No further investigation for this event is possible at this time.If additional information becomes available to indicate further evaluation is warranted, this record will be reopened.
|
| |
|
Search Alerts/Recalls
|