Model Number 403001000
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 12/04/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the defective device was not returned to the manufacturer and no lot number was available a simulation test was performed on a good sample.This simulation test confirmed that the reported problem - the distal tip of the bronchoscope being stuck in bend position - is most likely caused by excessive force being applied by the user to the control lever.The ascope 3 ifu states that the user should not exert excessive force to the control lever and bending section, as this may result in damage to the equipment.This mdr is resubmitted again as it was identified that original submission had not been successfully loaded into the fda database due to wrong file format.The initial mdr of this incident was submitted to the fda within the original reporting deadline.This submission represents a reload of data to ensure correct upload to the fda database.
 
Event or Problem Description
During an emergency bronchial fibroscopy in a (b)(6) year-old patient the distal end of the ascope 3 locked in a bent position.It was impossible to put it back in upright position.Patient outcome was not affected.However, the examination was interrupted because the ascope was difficult to direct.Extraction of the fiberscope from the intubation tube was complicated.There was a risk of extubation of the patient during the extraction of the fiberscope.No harm was caused to this patient, however, were this fault to recur it would pose a risk of causing serious injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASCOPE 3 REGULAR
Common Device Name
BRONCHOSCOPE (FLEXIBLE OR RIGID)
Manufacturer (Section D)
AMBU A/S
baltorpbakken 13
ballerup, dnk 2750
DA  2750
MDR Report Key11998704
Report Number1220828-2021-00001
Device Sequence Number3782479
Product Code EOQ
UDI-Device Identifier05707480121991
UDI-Public05707480121991
Combination Product (Y/N)N
Initial Reporter CountryFR
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Importer
Initial Reporter Occupation Other
Type of Report Initial
Report Date (Section B) 01/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model Number403001000
Device Catalogue Number403001000
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Event Location Hospital
Initial Date Received by Manufacturer Not provided
Initial Report FDA Received Date06/15/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No