| Model Number |
ETLW1610C146E |
| Medical Device Problem Code |
Defective Component (2292)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/27/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
An endurant ii stent graft system was attempted implanted in the patient for the endovascular treatment of a 62 mm abdominal aortic aneurysm.It was reported that during the index procedure, the device was inserted, where it was noted under fluoroscopy that the proximal markers of the device were not t lined up in the crimped state.It was stated that one marker was 3-4 mm lower that the other.The device was then removed from the patient and a etlw1610c156e was then used with no issues.As per the physician the cause of the event can not be determined.No additional clinical sequalae were reported and the patient is fine.
|
| |
|
Additional Manufacturer Narrative
|
|
Product analysis device decontaminated with cidex-opa pending further device testing to support the final product analysis findings.The device returned with the external slider in the home position.There was no deformation observed to the graft.Bunching was observed to the material at the proximal end of the graft.On deployment of the graft the material was no longer bunched.The proximal markers were confirmed to be on the same level around the circumference of the graft.The reported deviation was confirmed through analysis.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|