MEDTRONIC PUERTO RICO OPERATIONS CO. INTELLIS; STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR
|
Back to Search Results |
|
Model Number 97715 |
Device Problems
Electromagnetic Interference (1194); Energy Output Problem (1431); Overheating of Device (1437); Ambient Noise Problem (2877)
|
Patient Problems
Burning Sensation (2146); Paresthesia (4421); Insufficient Information (4580)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Concomitant medical products: product id: 977a260, serial#: (b)(4), product type: lead.Product id: 977a260, serial#: (b)(4), product type: lead.Product id: 977a260, serial/lot #: (b)(4), ubd: 11-dec-2018, udi#: (b)(4).Product id: 977a260, serial/lot #: (b)(4), ubd: 11-dec-2018, udi#: (b)(4).If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from a patient via manufacturer representative who was implanted with an implantable neurostimulator (ins) for spinal pain.It was reported that patient said during mri, there is a warm sensation running through leads.During ct and ultrasound patient complains of increased pulsing/tingling.No troubleshooting steps were taken.
|
|
Manufacturer Narrative
|
Continuation of d10: product id: 977a260, serial# (b)(6), product type: lead.Product id: 977a260, serial# (b)(6), product type: lead.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
The patient reported via a manufacturer representative warmth through out the body during mri scan, and during ultrasound and ct scan patient felt a stronger tingling in the lead.The manufacturer representative has checked the system's integrity and there is not an issue.
|
|