| Model Number |
TN,DXH 800HEMATOLOGY SYSTEM WITH FLOOR STAND |
| Medical Device Problem Codes |
Leak/Splash (1354); Blocked Connection (2888)
|
| Health Effect - Clinical Code |
Eye Injury (1845)
|
| Date of Event |
05/26/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The field service engineer (fse) reported that when removing a clog from the dxh800 instrument with a syringe, bleach splashed into the eye of an operator who was observing.The operator was wearing only a laboratory coat for ppe and was not wearing face protection at the time of the incident.The fse removed the clog from the instrument and the laboratory supervisor reminded the operator to use eye protection when partaking in troubleshooting activity.The operator did not sustain any long term or significant damage as a result of the splash in the eye.Bec internal identifier - case-(b)(4).
|
| |
|
Event or Problem Description
|
|
The field service engineer (fse) reported that while removing a clog from the dxh800 instrument with a syringe, bleach splashed into the eye of an operator who was observing.While troubleshooting the white blood cell (wbc) bath, the fse pushed bleach through the clogged lines with a syringe and tubing became loose and splashed bleach.An operator from the laboratory was observing and bleach splashed into their eyes.The operator then rinsed their eyes at the wash station and after that was taken to the emergency room.The operator was prescribed an eye ointment.The operator was out from work for two days but has since returned at full capacity and fully recovered from the incident.The operator was wearing only a laboratory coat for ppe and was not wearing face protection at the time of the incident.
|
| |
|
Search Alerts/Recalls
|