| Catalog Number |
8065751614 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/28/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Investigation, including root cause analysis, is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
|
| |
|
Event or Problem Description
|
|
A nurse reported trocar leakage during surgery.The cassette was replaced and the procedure was completed.There was no patient harm.
|
| |
|
Event or Problem Description
|
|
The event occurred during a vitrectomy surgery and the cassette pak was replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Three opened 23 gauge (ga) trocar assemblies in a small part tray (smpt) in a tray were received for the report of trocars leaking.The returned samples were visually inspected and were found to be non-conforming with damage to the septums.Leak testing could not be performed due to the damage of the sample.A review of the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria.The exact root cause for this complaint is unknown; the damaged condition of the valves could have contributed to the reported event; how and when the valve became damaged cannot be determined from the evaluation performed.The exact root cause for this complaint is unknown therefore specific action for this complaint cannot be taken.An acceptance quality inspection is performed to ensure product meets release acceptance criteria.This inspection includes evaluation for damaged valves.Complaints are reviewed and monitored at regular intervals to evaluate effectiveness of actions taken.No adverse trends have been observed associated with the reported product and event.No further actions are required.The manufacturer internal reference number is: (b)(4).
|
| |
|
Search Alerts/Recalls
|