MAUDE Adverse Event Report: BIOTRONIK SE & CO. KG ESTELLA DR-T; PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOTRONIK SE & CO. KG ESTELLA DR-T; PACEMAKER Back to Search Results
Model Number 371207
Medical Device Problem Code Battery Problem (2885)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/08/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was received for analysis.Prior to the analysis of the device, the quality documents accompanying the manufacturing process for this device were re-investigated and all production steps were performed accordingly.Particularly the final acceptance test proved the device functions to be as specified.Upon initial interrogation a warning message was triggered regarding the battery condition, confirming the clinical observation.The memory content of the device was inspected.The inspection revealed that the battery voltage decreased faster than expected.However, the current consumption was normal.The ability of the device to deliver therapies was verified.The anti-bradycardia pacing pulses proved to be normal and in amplitude and frequency as programmed in eri mode.There was no indication of a device malfunction.During subsequent analysis the device was opened and subjected to further investigation.The visual inspection of the inner assembly showed no anomalies.The measurement of the battery voltage confirmed an almost depleted battery.A thorough investigation of the electronic module did not show any abnormality.In particular, the current consumption was directly measured and proved to be normal and expected.The battery was sent to the manufacturer for further analysis.The manufacturing records of the battery were inspected, documenting that the battery parameters were within specification during the battery manufacturing.No anomalies were documented during the production process, associated with this battery.The battery was subjected to a visual and an electrical inspection as well as a microcalorimetry analysis.The visual inspection of the battery did not reveal any external signs of damage.The measurement of the battery voltage confirmed an almost depleted battery.The microcalorimetry analysis showed no anomalies.Next, the battery was opened for destructive analysis and the inner assembly was inspected.During analysis no internal electrical short was evident.However, the occurrence of a temporary short circuit that contributed to an increased internal self-depletion cannot be excluded.In summary, the root cause for the premature battery depletion could not be determined conclusively.Based on the analysis all therapy functions of the device were entirely available while the device was implanted and in service.
 
Event or Problem Description
It was reported that this device showed a battery error upon interrogation.Device was explanted and returned for analysis.The manufacturer analysis revealed a battery issue.No adverse patient events were reported, should additional information become available, this file will be updated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ESTELLA DR-T
Common Device Name
PACEMAKER
Manufacturer (Section D)
BIOTRONIK SE & CO. KG
woermannkehre 1
berlin 12359
DE  12359
MDR Report Key12002698
Report Number1028232-2021-03293
Device Sequence Number3560368
Product Code NVZ
UDI-Device Identifier04035479114299
UDI-Public04035479114299
Combination Product (Y/N)N
Initial Reporter CountryUS
PMA/510(K) Number
P950037
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2011
Device Explanted Year2020
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 06/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/30/2012
Device Model Number371207
Device Catalogue NumberSEE MODEL NO.
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/15/2020
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/10/2021
Initial Report FDA Received Date06/15/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization;
Patient Age87 YR
-
-