MAUDE Adverse Event Report: LAKE REGION MEDICAL ZIPWIRE HYDROPHILIC GUIDEWIRE; UROLOGICAL CATHETER AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LAKE REGION MEDICAL ZIPWIRE HYDROPHILIC GUIDEWIRE; UROLOGICAL CATHETER AND ACCESSORIES Back to Search Results
Model Number M006630208B1
Medical Device Problem Codes Fracture (1260); Device Damaged by Another Device (2915)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/15/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The device was not received for evaluation at the time of this report; therefore, no physical analysis of the device can be performed.A review of the device history records of the specimen device does not present any indication of manufacturing defect or anomaly that could have impacted on the event as reported.During manufacturing the production operators are instructed to 100% visually and tactilely inspect for any obvious defect, which includes a tactile examination of the entire length of each wire.In addition, during packaging of this product the operators are instructed to 100% visually inspect for any obvious defect prior to shipment.The history records indicate this product was final inspection tested at lake region medical and was determined to be acceptable.As noted in the device instructions for use (dfu) warnings, do not manipulate, advance and/or withdraw the zipwire hydrophilic guidewire through a metal cannula or needle.Manipulation, advancement, and/or withdrawal through a metal device may result in destruction and/or separation of the outer polymer jacket requiring retrieval.If a needle is used for initial placement, a plastic entry needle is recommended when using the zipwire hydrophilic guidewire.Extreme caution should be observed when used with a one-wall puncture needle.Use extreme caution when using a laser, making sure to avoid contact with the zipwire hydrophilic guidewire.Direct contact may cause damage to the wire and/or sever the wire.Manipulate the zipwire hydrophilic guidewire slowly and carefully in the urinary system while confirming the behavior and location of the wire's tip under fluoroscopy.Excessive manipulation of the zipwire hydrophilic guidewire without fluoroscopic confirmation may result in perforation or trauma of the linings or associated tissues, channels or ducts.If any resistance is felt or if the tip's behavior and/or location seem improper, stop manipulating the wire and/or catheter and determine the cause by fluoroscopy.Failure to exercise proper caution may result in bending, kinking, separation of the zipwire hydrophilic guidewire's tip, damage to the catheter, or damage to the urinary system.If necessary, remove the zipwire hydrophilic guidewire and ancillary device or scope as a complete unit to avoid complications.The dfu precautions also indicate, the zipwire hydrophilic guidewire should be advanced through the scope using short, deliberate 2-3cm movements to prevent inadvertent damage to the device or patient.· due to variations in certain catheter tip diameters, abrasion of the hydrophilic coating may occur during manipulation.If any resistance is felt during introduction of the catheter, it is advisable to stop using such catheters.The complaint is non-verifiable as the product was not returned for evaluation.At this time, it is not possible to assign a definitive root cause for the event as reported.If any further information is received, a follow up medwatch report will be submitted.
 
Event or Problem Description
Event description: per medwatch, it was reported that: zipwire general-purpose non-vascular guidewire was being used to guide into the patient right ureter for a stent exchange and lithotripsy.The zipwire guidewire was found to be severed, confirmed through fluoroscopy.The guidewire was retrieved immediately by the surgeon, and the stent was placed as planned.No harm to the patient was identified, only extending the length of the surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ZIPWIRE HYDROPHILIC GUIDEWIRE
Common Device Name
UROLOGICAL CATHETER AND ACCESSORIES
Manufacturer (Section D)
LAKE REGION MEDICAL
340 lake hazeltine drive
chaska MN 55318
MDR Report Key12003943
Report Number2126666-2021-00018
Device Sequence Number14279194
Product Code EYA
UDI-Device Identifier08714729764267
UDI-Public08714729764267
Combination Product (Y/N)N
Initial Reporter StateAZ
Initial Reporter CountryUS
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Risk Manager
Type of Report Initial
Report Date (Section B) 06/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/31/2023
Device Model NumberM006630208B1
Device Catalogue Number630-208B
Device Lot Number5975145
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/01/2021
Initial Report FDA Received Date06/15/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age82 YR
-
-