• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AESCULAP AG PAS-PORT PROXIMAL ANASTOMOSIS SYSTEM; ANASTOMOSIS DEVICES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AESCULAP AG PAS-PORT PROXIMAL ANASTOMOSIS SYSTEM; ANASTOMOSIS DEVICES Back to Search Results
Model Number FC700SU
Medical Device Problem Code Product Quality Problem (1506)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/14/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Manufacturer evaluation: investigation on-going.Should relevant additional information / investigation results become available, a supplemental medwatch report will be submitted.
 
Event or Problem Description
It was reported to aesculap ag that a pas-port proximal anastomosis system (part # fc700su) was used during a procedure performed on (b)(6) 2021.According to the complainant, during the surgery, the staples were deployed outside of the aorta.After the anastomosis, the staples was deployed outside the aorta and only left a hole in the aorta.A side clamp was used on the aorta and the hole was closed by hand sewing.It is not currently known how the case was completed.The nurse threw away the pps body with the saphenous vein graft (svg) inside.An additional medical intervention was necessary.Although requested, additional information has not been made available.The adverse event is filed under aag reference (b)(4).
 
Additional Manufacturer Narrative
Investigation results: as of the date of this report the complaint product was not provided for investigation.Therefore, a thorough investigation is not possible.Batch history review: the device quality and manufacturing history records have been checked for the available lot number and were found to be according to our specifications valid at the time of production.Review of the complaint history revealed that 5 similar complaints have been filed against products from this batch number.The review of risk assessment revealed that the overall risk level (severity 4(5) x probability of occurrence 2(5)) according to din en iso 14971 is still acceptable.Conclusion and measures / preventive measures: based upon the investigation results a clear root cause conclusion cannot be drawn.There is no indication for a material-, manufacturing- or design-related failure.In the event that the complaint product will be provided for investigation in the future, an update of this report will be provided unsolicited.Based upon the investigations results a product safety case (b)(4) was initiated.Any action regarding capa will be addressed with this case.A product recall (b)(4) was initiated.Product not sold nor marketed any longer.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PAS-PORT PROXIMAL ANASTOMOSIS SYSTEM
Common Device Name
ANASTOMOSIS DEVICES
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
MDR Report Key12004473
Report Number9610612-2021-00438
Device Sequence Number14342514
Product Code FZP
UDI-Device Identifier04046955241026
UDI-Public4046955241026
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K202124
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,distri
Initial Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date (Section B) 10/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date05/24/2022
Device Model NumberFC700SU
Device Catalogue NumberFC700SU
Device Lot Number52632015
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/20/2021
Supplement Date Received by Manufacturer09/28/2021
Initial Report FDA Received Date06/15/2021
Supplement Report FDA Received Date10/28/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-