Model Number EM100-A
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 03/31/2019
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Report confirmed.The device was tested and the max temperature was measured to be 140°f.The likely cause was identified as worn front and rear bearings due to inadequate preventative maintenance.The device user manual warnings section includes instructions that heavy side loads and/or long operating times may cause the device to overheat.If the device overheats, discontinue use and rest the device by using it intermittently, or wrap the device in a moist sterile towel.Use of an overheated device may cause injury to the patient or operator.The preventive maintenance/service manual for the legend system specifies service intervals for devices based on the hospital usage level.The maximum specified service interval is 24 months.Device has been in use for approximately 121 months with no record of factory service during this period.We will continue to track and trend this complaint type.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Repair request initiated for device with the report of overheating.No patient impact reported.Repair request escalated to a product event based on reason for return and due to customer indication that this report is a complaint.On follow up, it was reported that there was no patient or staff impact.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MOTOR LEGEND EHS
Common Device Name
MOTOR, DRILL, ELECTRIC
Manufacturer (Section D)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
MDR Report Key12005858
Report Number1625507-2021-00149
Device Sequence Number11589186
Product Code HBC
UDI-Device Identifier10885074106486
UDI-Public10885074106486
Combination Product (Y/N)N
Initial Reporter CountryTC
PMA/510(K) Number
K012456
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 06/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberEM100-A
Device Catalogue NumberEM100-A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/18/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/26/2019
Initial Report FDA Received Date06/15/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-