MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 EQUISTREAM XK HEMODIALYSIS CATH W/STYLET, ST, AIRGUARD, 16 FR. 23CM; DIALYSIS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 EQUISTREAM XK HEMODIALYSIS CATH W/STYLET, ST, AIRGUARD, 16 FR. 23CM; DIALYSIS CATHETER Back to Search Results
Model Number 5913230
Medical Device Problem Code Contamination /Decontamination Problem (2895)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/10/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history record is currently being performed.The return of the sample is pending.The investigation of the reported event is currently underway.Expiry date: 08/2022.
 
Event or Problem Description
It was reported that during catheter placement procedure, a small fragment was allegedly found in one of the hubs while aspirating.It was further reported that the fragment did not enter the patients body.There was no reported patient injury.
 
Event or Problem Description
It was reported that during dialysis catheter placement procedure in the right chest , a small fragment was allegedly found in one of the hubs while aspirating.It was further reported that the fragment did not enter the patients body.There was no reported patient injury.
 
Additional Manufacturer Narrative
H10: manufacturing review: the device history records have been reviewed and this lot met all release criteria.There was nothing found to indicate there was a manufacturing related cause for this event.Investigation summary: the physical device was not returned for evaluation.No photos were provided for review.Therefore, the investigation is inconclusive for the reported foreign material as no objective evidence was provided for review.A definitive root cause could not be determined based upon the available information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 08/2022).H11: section a through f ¿ the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EQUISTREAM XK HEMODIALYSIS CATH W/STYLET, ST, AIRGUARD, 16 FR. 23CM
Common Device Name
DIALYSIS CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
MDR Report Key12006433
Report Number3006260740-2021-02298
Device Sequence Number14278874
Product Code MSD
UDI-Device Identifier00801741013843
UDI-Public(01)00801741013843
Combination Product (Y/N)N
Initial Reporter StateNC
Initial Reporter CountryUS
PMA/510(K) Number
K090101
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number5913230
Device Catalogue Number5913230
Device Lot NumberREEZ3608
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/21/2021
Supplement Date Received by Manufacturer09/30/2021
Initial Report FDA Received Date06/16/2021
Supplement Report FDA Received Date10/05/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
-
-