| Model Number |
2900163 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/24/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Information was received from multiple sources (manufacturer representative, healthcare provider) regarding a device used for discet omy at the l3-4, l4-5 levels.Pre-operative diagnosis : l3-5 lumbar stenosis it was reported that doctor was performing a discectomy at the l3-4, l4-5 levels and prepping to implant a capstone device.Upon using the uterine curette, he was scrapping the endplates with the instrument and felt it break loose.He removed the instrument and noticed a crack on one side of the curette.He immediately placed it off the sterile field.No fragment was left inside the patient's body.There were no patient symptoms or complications as a result of this event.No further complications were reported regarding the event.Update : the product came in contact with the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
H3: product analysis :part # 2900163; lot # im13e006 visual and optical inspection confirmed the tip of the curette has been cracked.This type of damage is consistent with bend stress overload.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|