| Model Number |
RFDE-SI |
| Medical Device Problem Code |
Impedance Problem (2950)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
12/23/2020
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
One simplicity radiofrequency electrode was received for evaluation.The device met specifications for temperature response and no short circuits were detected.No functional anomalies were noted; however the tubing was bent and electrode 3 was fractured but still attached.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported fracture remains unknown.
|
| |
|
Event or Problem Description
|
|
This report is to advise of an event noted during analysis which revealed a fractured electrode.
|
| |
|