Model Number SPL-T
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Upon inspection of the returned device, the suction port was found to be broken off.No additional defects or failures were noted.This device is a non-repairable and was returned to the customer unrepaired.The device has been previously serviced by olympus.Therefore, a service history review will replace dhr review.All previous inspection records show that no problems were found to the device except for minor scratches on the housing.A definitive root cause of the found defect could not be determined.However, damage to the suction port is often incurred as a result of mishandling.The device ifu states "this product is a precision device; handle it with care.Avoid rough or violent handling, which may cause equipment damage," (page 1).Olympus will continue to monitor complaints for this device through regular trending activities and take action if triggered by osta qms procedure.
 
Event or Problem Description
This is an asset return which was found to have a broken suction port during estimation.There was no device issue or patient injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SHOCKPULSE LITHOTRIPSY TRANSDUCER.
Common Device Name
LITHOTRIPTOR, ULTRASONIC
Manufacturer (Section D)
GYRUS ACMI, INC
136 turnpike road
southborough PA 01772
MDR Report Key12008955
Report Number8010047-2021-07599
Device Sequence Number3197620
Product Code FEO
UDI-Device Identifier00855279005023
UDI-Public00855279005023
Combination Product (Y/N)N
Initial Reporter StateCA
Initial Reporter CountryUS
PMA/510(K) Number
K142428
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date (Section B) 06/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSPL-T
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/25/2021
Was the Report Sent to FDA? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/26/2021
Initial Report FDA Received Date06/16/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-