Model Number 1570-11-100
Medical Device Problem Code Degraded (1153)
Health Effect - Clinical Codes Foreign Body Reaction (1868); Hypersensitivity/Allergic reaction (1907); Pain (1994); Discomfort (2330); Unspecified Tissue Injury (4559); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 10/26/2020
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).Initial reporter occupation: lawyer.(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Asr claim received.After review of medical record, patient was revised to address right hip metallosis with soft tissue reaction.Patient experienced pain and discomfort and elevated metal ion.Operative findings stated that there was significant corrosion at the head and neck junction of femoral stem.There were sign of pseudotumor.Revision notes stated that synovial fluid was obtain for culture.Doi: (b)(6) 2007; dor: (b)(6) 2020; right hip.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SUMMIT POR TAPER SZ4 HI OFF
Common Device Name
SUMMIT HIP STEM : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key12009003
Report Number1818910-2021-12768
Device Sequence Number3425956
Product Code LPH
UDI-Device Identifier10603295060055
UDI-Public10603295060055
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
PMA/510(K) Number
K001991
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2002
Device Explanted Year2020
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 01/24/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number1570-11-100
Device Catalogue Number157011100
Device Lot NumberBW6AD1000
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/08/2021
Supplement Date Received by Manufacturer06/18/2021
Initial Report FDA Received Date06/16/2021
Supplement Report FDA Received Date06/22/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ADAPTER SLEEVES 12/14 +2; ASR ACETABULAR CUPS 54; ASR UNI FEMORAL IMPL SIZE 47; SUMMIT POR TAPER SZ4 HI OFF; ADAPTER SLEEVES 12/14 +2; ASR ACETABULAR CUPS 54; ASR UNI FEMORAL IMPL SIZE 47; SUMMIT POR TAPER SZ4 HI OFF
Outcome Attributed to Adverse Event Required Intervention;
Patient Age58 YR
-
-