• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHYSIO-CONTROL, INC. - 3015876 LIFEPAK 15 DEFIBRILLATOR/MONITOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHYSIO-CONTROL, INC. - 3015876 LIFEPAK 15 DEFIBRILLATOR/MONITOR; AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE) Back to Search Results
Model Number 15
Medical Device Problem Code Failure to Charge (1085)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/20/2020
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Physio-control evaluated the customer's device and verified that the event code was logged in the device's memory; however, the issue could not be duplicated during testing.After completing other unrelated repairs, proper device operation was observed through functional and performance testing and the device was returned to the customer for use.The cause of the reported issue could not be determined.
 
Event or Problem Description
The customer contacted physio-control to report that they observed an event code logged in the device's memory.The event code logged is indicative of a failure of the device to charge for defibrillation energy.As a result, defibrillation therapy may not be available, if it was needed.There was no report of patient use associated with the reported event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEPAK 15 DEFIBRILLATOR/MONITOR
Common Device Name
AUTOMATED EXTERNAL DEFIBRILLATORS (NON-WEARABLE)
Manufacturer (Section D)
PHYSIO-CONTROL, INC. - 3015876
11811 willows road ne
redmond WA 98052
MDR Report Key12009945
Report Number0003015876-2021-01245
Device Sequence Number11842878
Product Code MKJ
UDI-Device Identifier00883873871836
UDI-Public00883873871836
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K142430
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 06/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number15
Device Catalogue Number99577-001943
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/20/2021
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/20/2021
Initial Report FDA Received Date06/16/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-