DEPUY ORTHOPAEDICS INC US AML LG STATURE 16.5MM; AML IMPLANT : HIP FEMORAL STEM Back to Search Results
Model Number 1554-02-165
Medical Device Problem Codes Inadequacy of Device Shape and/or Size (1583); Osseointegration Problem (3003); Migration (4003)
Health Effect - Clinical Codes Pain (1994); Inadequate Osseointegration (2646)
Date of Event 06/02/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the patient was experiencing a lot of pain.Stem subsided, surgeon believed it to be infected but it was not.Stem looked undersized on original xray.Liner swapped and larger stem implanted with new head.Original size was 16.5 lg, implanted a 19.5 lg.It was also indicated that there was loosening of the stem at the bone to implant interface.Doi: (b)(6) 2018; dor: (b)(6) 2021; affected side: left hip.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed, and the investigation will be re-opened as necessary.Device history lot : the product investigation found no evidence suspecting an error in the material, manufacturing, inspection or sterile processing that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AML LG STATURE 16.5MM
Common Device Name
AML IMPLANT : HIP FEMORAL STEM
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key12010164
Report Number1818910-2021-12815
Device Sequence Number3304035
Product Code LPH
UDI-Device Identifier10603295058380
UDI-Public10603295058380
Combination Product (Y/N)N
Initial Reporter StateIN
Initial Reporter CountryUS
PMA/510(K) Number
K012364
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1554-02-165
Device Catalogue Number155402165
Device Lot Number293969
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/02/2021
Supplement Date Received by Manufacturer07/13/2021
Initial Report FDA Received Date06/16/2021
Supplement Report FDA Received Date07/14/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age72 YR
-
-