| Model Number |
HPT040010A |
| Medical Device Problem Code |
Complete Blockage (1094)
|
| Health Effect - Clinical Codes |
Bacterial Infection (1735); Obstruction/Occlusion (2422); Thrombosis/Thrombus (4440)
|
| Date of Event |
11/13/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Cbas® heparin surface incorporates cbas-heparin manufactured from heparin sodium api, which is covalently bound to the device surface and is essentially non-eluting.
|
| |
|
Event or Problem Description
|
|
The following was reported to gore: on (b)(6) 2020, a patient underwent a right thoracotomy as part of a blalock-taussig shunt procedure utilizing a gore® propaten® vascular graft configured for pediatric shunt device.Physician confirmed all shunts are on prophylactic heparin until they¿re able to transition to aspirin and that aspirin responsiveness is checked in everyone prior to stopping heparin.On the same day as the initial procedure, (b)(6) 2020, evidence of decreasing saturations and echo studies could not confirm shunt patency.A few hours after, no shunt flow could be detected.On the following day, (b)(6) 2020, the shunt is difficult to see by doppler and there doesn't appear to be any flow through it.A thrombectomy was performed and the original shunt remained.Progressive narrowing of the shunt was described with echo beginning on (b)(6), 2021.On (b)(6) 2021 shunt ligation and takedown was performed and purulent material was found around the shunt, which was subtotally thrombosed.Pcr detected staphylococcus epidermidis around the shunt.
|
| |
|
Additional Manufacturer Narrative
|
|
H6: type of investigation, investigation findings, and investigation conclusions updated to reflect completion of investigation.
|
| |
|
Additional Manufacturer Narrative
|
|
Completed sterilization records review: a review of the manufacturing records indicated the lot met all pre-release sterilization specifications.
|
| |
|