• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. STRAIGHT SCREWDRIVER; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. STRAIGHT SCREWDRIVER; PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Model Number 71362293
Medical Device Problem Code Material Deformation (2976)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/01/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that, during thr surgery, a straight screwdriver was noticed to be stripped.No significant surgical delay was reported to occur due to this issue, as a smith and nephew backup device was available.Neither patient injury nor other complications were reported.
 
Additional Manufacturer Narrative
The device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.A complaint history review found related failures for the listed device; this failure mode will be monitored for future complaints for any necessary corrective actions.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
STRAIGHT SCREWDRIVER
Common Device Name
PROSTHESIS, KNEE, FEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12012554
Report Number1020279-2021-05202
Device Sequence Number4408913
Product Code HRY
UDI-Device Identifier03596010454188
UDI-Public03596010454188
Combination Product (Y/N)N
Initial Reporter StateSC
Initial Reporter CountryUS
PMA/510(K) Number
K121393
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71362293
Device Catalogue Number71362293
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 06/01/2021
Supplement Date Received by Manufacturer07/29/2021
Initial Report FDA Received Date06/16/2021
Supplement Report FDA Received Date08/03/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-