| Model Number |
151.053 |
| Health Effect - Clinical Code |
Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
|
| Date of Event |
06/03/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Neither the device nor any imaging was available for evaluation.The cause of the reported issue could not yet be determined.
|
| |
|
Event or Problem Description
|
|
It was reported by a representative from the uk that a revision surgery was done for a fortify cage that collapsed post-operatively, causing adverse effects.
|
| |
|
Event or Problem Description
|
|
It was reported by a representative from the uk that a revision surgery was done for a fortify cage that collapsed post-operatively, causing adverse effects.
|
| |
|
Additional Manufacturer Narrative
|
|
The device was returned for evaluation.A micro ct scan revealed nothing out of the ordinary except for excessive wear on the gear tooth at the point of contact with the locking gear.However, an exact cause of the reported issue cannot be determined.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information provided that the surgeon inserted the screws into c3, the implant moved at c6 resulting in the surgeon not placing screws in the inferior endplate.Imaging provided showed a loss of height in the core of 3 threads or approximately 4.5mm.The cause of the reported issue could not yet be determined.The following sections were updated: b4, b6, g6, h6 and h10.
|
| |
|
Event or Problem Description
|
|
It was reported by a representative from the uk that a revision surgery was done for a fortify cage that collapsed post-operatively, causing adverse effects.
|
| |
|