| Model Number |
120-650 |
| Medical Device Problem Code |
Device Appears to Trigger Rejection (1524)
|
| Health Effect - Clinical Code |
Capsular Contracture (1761)
|
| Date of Event |
09/15/2020
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
A review of the device history record has been completed.No deviations or non-conformances noted.The event of "capsular contracture unknown baker grade" is a physiological complication and analysis of the device generally does not assist allergan in determining a probable cause for this event.Further information from the reporter regarding event, product, or patient details has been requested.No additional information is available at this time.Reason for reoperation: capsular contracture unknown baker grade.
|
| |
|
Event or Problem Description
|
|
Patient reported left side pain with capsular contracture, baker grade unknown.The device remains implanted.
|
| |
|
Event or Problem Description
|
|
Healthcare professional reported left side capsular contracture as baker grade iv.Device has been explanted and replaced.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluation: visual analysis of the returned device identified, crease fold, wear abrasion, deformation yellow particles on the shell, and overweight (even though the device is currently overweight, it is not considered a workmanship since all units of the weight report attached are within specification when released in the manufacturing process).After autoclave cycle was observed: cloudy color in the gel.Based on the device analysis the final assessment is: no issues found related with the manufacturing process.
|
| |
|
Event or Problem Description
|
|
Patient reported left side pain with capsular contracture, baker grade unknown.Healthcare professional reported left side capsular contracture as baker grade iv.Device has been explanted and replaced.
|
| |
|