| Model Number |
ESP100-21-00 |
| Medical Device Problem Codes |
Obstruction of Flow (2423); Biocompatibility (2886)
|
| Health Effect - Clinical Codes |
Dyspnea (1816); Thrombosis/Thrombus (4440); Heart Failure/Congestive Heart Failure (4446)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The results/method and conclusion codes along with investigation results will be provided in the final report.
|
| |
|
Event or Problem Description
|
|
It was reported that an (b)(6) year old patient implanted with an epic supra valve w/flexfit on (b)(6) 2021, presented with symptoms of congestive heart failure (chf) reporting dyspnea, regurgitation, and high gradients during a follow-up appointment in the clinic.A clot was confirmed on the valve.At the discovery of the thrombus, the patient was not on any anticoagulant or antiplatelet therapy.The patient was put on heparin and then switched to coumadin.The physician reported the valve remained implanted.The patients mean gradient has decreased into the 20s.No additional information was provided.
|
| |
|
Additional Manufacturer Narrative
|
|
An event of thrombus was reported.A more comprehensive assessment could not be performed as the device remains implanted was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications.Based on the information received, the cause of the reported incident could not be conclusively determined.
|
| |
|