| Model Number |
PXSLIMSTR-B |
| Medical Device Problem Codes |
Break (1069); Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/19/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.
|
| |
|
Event or Problem Description
|
|
The patient was undergoing a medical procedure using a px slim delivery microcatheter (px slim).During the procedure, the physician experienced resistance while advancing the px slim.Therefore, the px slim was removed.Upon removal, it was noted that the proximal end of the px slim was broken.The procedure was completed using a new px slim.There was no report of an adverse effect to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation of the returned px slim confirmed fractures in the proximal shaft.If the device is forcefully manipulated against resistance during use, damage such as kinks and subsequent fractures may occur.The root cause of resistance during the procedure could not be determined.Further evaluation revealed that the device was kinked throughout its length and the strain relief was stretched.This damage was likely incidental to the complaint.Penumbra catheters are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
| |
|