PENUMBRA, INC. LANTERN DELIVERY MICROCATHETER; DQY Back to Search Results
Model Number PXSLIMLAN115T45
Medical Device Problem Code Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/18/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report numbers: 3005168196-2021-01348, 3005168196-2021-01349, 3005168196-2021-01351.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the splenic artery using pod packing coils (pod pcs) and lantern delivery microcatheters (lanterns).During the procedure, a pod pc was determined to be too long for the coil pack.When retracting the pod pc, the physician experienced resistance, and the pod pc unintentionally detached inside the lantern.The detached pod pc and lantern were then removed together.Subsequently, the physician was unable to flush the pod pc out of the lantern and decided to use a new lantern and new pod pc.The pod pc was also determined to be too long for the coil pack.When retracting the pod pc, the physician experienced resistance, and the pod pc unintentionally detached inside the lantern.The detached pod pc and lantern were then removed together.No additional information regarding the completion of the procedure was provided.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LANTERN DELIVERY MICROCATHETER
Common Device Name
DQY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key12014237
Report Number3005168196-2021-01350
Device Sequence Number19176562
Product Code DQY
UDI-Device Identifier00814548016627
UDI-Public00814548016627
Combination Product (Y/N)Y
Initial Reporter StateTN
Initial Reporter CountryUS
PMA/510(K) Number
K152840
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/16/2021
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date12/04/2023
Device Model NumberPXSLIMLAN115T45
Device Catalogue NumberPXSLIMLAN115T45
Device Lot NumberF102496
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 05/18/2021
Initial Report FDA Received Date06/16/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age42 YR
Date Report Sent to FDA01/01/2005
-
-