INTERVASCULAR SAS INTERGARD WOVEN AORTIC ARCH; PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER Back to Search Results
Model Number HEWAA2610080810/1
Medical Device Problem Code Manufacturing, Packaging or Shipping Problem (2975)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/12/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
It was reported that the product is available for investigation; it is in the process of being returned to intervascular for examination.The device history records review is in progress, conclusion is pending.The review of historical data indicated that one other similar complaint, for yellow stains on the inner lid, was reported for the same type of product and same sterilization lot number.For this older complaint, the conducted investigation suggested that the device was not defective at the time of manufacturing.However, the product, as it was returned to the manufacturer, was confirmed as not conform to the specification.The root cause was unclear but probably related to inappropriate storage conditions by logistical actors.Please note that, according to manufacturing records of both involved products, coating was not performed the same day for both products, and the coating dispersions used were different.One retention sample from another lot was selected, based on the type of product and the coating date.A visual inspection by the qa department will be performed.The investigation is still ongoing.A follow-up report will be sent upon completion of the investigation.
 
Event or Problem Description
The physicians found the outer box worn and there are yellowish stains on the inner layer of the cover of tray.The product package does not meet the hospital¿s requirements.The issue was found during inspection, before surgery.The product has not been used on a patient.
 
Event or Problem Description
See initial mfr report #: 1640201-2021-00020.Complaint # (b)(4).
 
Additional Manufacturer Narrative
(3331/213) the device history records review concluded that there is no non-conformance / planned deviation in relation with the event reported.(4102/213) the selected retention sample was inspected by the qa supervisor.His conclusions are as follows:"the inspection of the retention product including the control of the packaging components has been completed.We can confirm that on the inner and outer primary packaging components no stain/halo is apparent that would be related to complaint # (b)(4).This retention sample is compliant from a product/packaging perspective." (10/3233) the defective product has been received by intervascular.The examination is pending.(11) the investigation is still ongoing.A follow-up report will be sent upon completion of the investigation.
 
Additional Manufacturer Narrative
(10/111) the involved product was returned to intervascular and was visually inspected by our quality assurance (qa) supervisor.The result are as follows: "there are several yellowish/orange stains on the inner lid, but the outer lid and the box look normal, with no visible stains of color or otherwise on the product." "the products, being non-conforming and coming from a customer, are physically rejected.¿ the product is not compliant with our list of standards for acceptation and rejection for packaging.Following the visual inspection a non conformity report has been initiated in order to identify the root cause.It concludes that root cause could be related to packaging design inducing a prolonged contact between product and tyvek lid resulting in glycerol stains on the lid.Please not that glycerol is an integral part of the product.(67) the outcome of the investigation which included all available information and the inspection performed on the returned product would tend to indicate that the issue was related to packaging design inducing a prolonged contact between product and internal tyvek lid resulting in glycerol stains on the lid.However, the observed glycerol traces are not an unusual phenomenon and do not affect the product quality and performance and constitutes a cosmetic defect.
 
Event or Problem Description
See initial mfg report #: 1640201-2021-00020.Complaint # (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERGARD WOVEN AORTIC ARCH
Common Device Name
PROSTHESIS, VASCULAR GRAFT, OF 6MM AND GREATER DIAMETER
Manufacturer (Section D)
INTERVASCULAR SAS
zone industrielle athelia i
la ciotat 13705
FR  13705
MDR Report Key12016244
Report Number1640201-2021-00020
Device Sequence Number19689467
Product Code DSY
UDI-Device Identifier00384401010746
UDI-Public(01)00384401010746
Combination Product (Y/N)N
Initial Reporter CountryTW
PMA/510(K) Number
K013651
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source distributor
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 08/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/31/2022
Device Model NumberHEWAA2610080810/1
Device Catalogue NumberHEWAA2610080810/1
Device Lot Number17J21
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/07/2021
Initial Date Received by Manufacturer 05/24/2021
Supplement Date Received by Manufacturer06/25/2021
07/22/2021
Initial Report FDA Received Date06/17/2021
Supplement Report FDA Received Date07/20/2021
08/10/2021
Patient Sequence Number1
-
-