| Model Number |
4965-15 |
| Medical Device Problem Codes |
Electrical /Electronic Property Problem (1198); Low impedance (2285); High Capture Threshold (3266)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/25/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Concomitant medical products: w1dr01 ipg, implanted: (b)(6) 2018.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
It was reported that the right atrial (ra) lead exhibited high thresholds, atrial lead position check fail and low unipolar impedance.It was also reported that the right ventricular (rv) lead exhibited low unipolar impedance, high thresholds, short v-v intervals.The leads were explanted and replaced.No patient complications have been reported as a result of this event.
|
| |
|
Additional Manufacturer Narrative
|
|
Product event summary: the lead was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory indicated the impedance of the atrial pacing lead was below the expected lower range.Analysis of the device memory indicated atrial pacing capture threshold was elevated.The device memory indicated atrial lead position check failed.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|