• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. GEN II 7.5MM RESUR PAT 29MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. GEN II 7.5MM RESUR PAT 29MM; PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL Back to Search Results
Catalog Number 71932635
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Date of Event 05/28/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that, after total knee arthroplasty had been performed on (b)(6) 2020, the patient experienced an infection.This adverse event was solved by performing a revision surgery on (b)(6) 2021.The reported devices were a journey ii bcs xlpe articular insert size 3-4 lt 9mm, a journey tibia base np lt size 3 , a genesis ii 7.5mm resur patella 29mm and a jrny ii bcs fem cocr lt sz 4.The surgeon stated the devices were not defective.
 
Additional Manufacturer Narrative
The device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.The clinical / medical investigation concluded that, per complaint details, a revision was performed due to infection approximately 7.5 months post implantation.Reportedly, the surgeon stated the devices were not defective.It was communicated that the requested clinical documentation is not available for inclusion in the medical investigation.The clinical root cause was reportedly infection; however, the root cause of the infection remains unknown.The patient impact beyond the reported infection and revision could not be determined.No further medical assessment could be rendered at this time.Should clinically relevant documentation become available in the future, this case may be re-opened/re-assessed.A review of complaint history did not reveal additional complaints for the listed batch.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of the risk management file and instructions for use documents revealed this failure mode was previously identified.A review of the sterilization records revealed the batch was sterilized within normal parameters.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Possible causes could include but not limited to contamination, patient reaction, and post-operative healing issue.Based on this investigation, the need for corrective action is not indicated.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GEN II 7.5MM RESUR PAT 29MM
Common Device Name
PROSTHESIS, KNEE, PATELLOFEMOROTIBIAL, SEMI-CONSTRAINED, CEMENTED, POLYMER/METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12017303
Report Number1020279-2021-05207
Device Sequence Number11663783
Product Code JWH
UDI-Device Identifier00885556056516
UDI-Public885556056516
Combination Product (Y/N)N
Initial Reporter CountryJA
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2020
Device Explanted Year2021
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Catalogue Number71932635
Device Lot Number20EM11736
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/28/2021
Supplement Date Received by Manufacturer07/28/2021
Initial Report FDA Received Date06/17/2021
Supplement Report FDA Received Date07/29/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Other;
Patient Age70 YR
-
-