• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: CAYENNE MEDICAL QUATTRO SUTURE PASSER NEEDLE; NEEDLE, SUTURING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CAYENNE MEDICAL QUATTRO SUTURE PASSER NEEDLE; NEEDLE, SUTURING Back to Search Results
Model Number N/A
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/12/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Foreign: (b)(6).Customer has not indicated that the product will be returned to zimmer biomet for investigation.If device returns or once the investigation has been completed, a follow-up mdr will be submitted.
 
Event or Problem Description
It was reported during a right shoulder arthroscopy cuff repair operation, the hook end (tip end) from the quattro suture passer was missing.Attempts were made to find the missing tip but were not found, it was reported to not be in the joint and most likely to be the cuff, soft tissue or outside of the patient.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional and/or corrected information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Medical records were not provided.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUATTRO SUTURE PASSER NEEDLE
Common Device Name
NEEDLE, SUTURING
Manufacturer (Section D)
CAYENNE MEDICAL
16597 n 92nd street
scottsdale AZ 85260
MDR Report Key12017310
Report Number3006108336-2021-00025
Device Sequence Number3604504
Product Code GAB
UDI-Device Identifier00887868270147
UDI-Public(01)00887868270147(17)230825(10)75957-1
Combination Product (Y/N)N
Initial Reporter CountryUK
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source distributor,foreign
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 08/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date08/25/2023
Device Model NumberN/A
Device Catalogue NumberCM-9011
Device Lot Number75957-1
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 05/26/2021
Supplement Date Received by Manufacturer08/13/2021
Initial Report FDA Received Date06/17/2021
Supplement Report FDA Received Date08/20/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
-
-