| Model Number |
1806-0050S |
| Medical Device Problem Codes |
Fracture (1260); Improper or Incorrect Procedure or Method (2017)
|
| Health Effect - Clinical Code |
Foreign Body In Patient (2687)
|
| Date of Event |
05/07/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Upon completion of investigation, additional information will be provided in a supplemental report.
|
| |
|
Event or Problem Description
|
|
As reported: "in the procedure of the t2 femur nail, due to the complexity of the fracture, guide pins are passed, and the guide needle is passed 18060050s for the placement of the screw, the resident cuts the pins that have been passed to reduce the fracture and between them cut the guide needle through error, so the surgeon decided to leave the guide needle to help in reducing the fracture.".
|
| |
|
Additional Manufacturer Narrative
|
|
D9 / h3 updated to indicate device was not returned.The reported event could not be confirmed, since the device was not returned for evaluation and no other evidences were provided.The device inspection was not possible as the product was not returned for investigation.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.The operative technique instructs user that: ¿while removing the power tool the guide wire pusher will keep the guide wire in place.Remove the guide wire prior to drilling and inserting the locking screws.¿ more detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.If device is returned or any further information is provided, the investigation report will be reassessed.
|
| |
|
Event or Problem Description
|
|
As reported: "in the procedure of the t2 femur nail, due to the complexity of the fracture, guide pins are passed, and the guide needle is passed 18060050s for the placement of the screw, the resident cuts the pins that have been passed to reduce the fracture and between them cut the guide needle through error, so the surgeon decided to leave the guide needle to help in reducing the fracture.".
|
| |
|
Search Alerts/Recalls
|