Model Number DL-330-10-K
Medical Device Problem Code Impedance Problem (2950)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/18/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Neuropace evaluated the data uploaded to pdms (including ecogs and impedances).The review of the data was consistent with a potential lead break.The explanted product was discarded by the treating center and was not returned to neuropace for analysis.
 
Event or Problem Description
The treating clinician performed a review of the patient's device data and identified abnormal impedances and artifact present on the recorded ecogs on (b)(6) 2020.Responsive stimulation was disabled on contact 2 of the affected lead.The patient's response was monitored until such time the clinician deemed a replacement of the affected lead was necessary.A lead replacement procedure was performed on (b)(6) 2021 without complication.There is no clear indication of head trauma or other inciting event.No other information was reported from the treating center.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NEUROPACE RNS SYSTEM
Common Device Name
NEUROPACE RNS SYSTEM
Manufacturer (Section D)
NEUROPACE, INC.
455 n. bernardo ave.
mountain view CA 94043
MDR Report Key12018326
Report Number3004426659-2021-00031
Device Sequence Number11619394
Product Code PFN
UDI-Device Identifier00855547005045
UDI-Public010085554700504517210213
Combination Product (Y/N)N
Initial Reporter StateMI
Initial Reporter CountryUS
PMA/510(K) Number
P100026
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2018
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device No Information
Device Model NumberDL-330-10-K
Device Catalogue Number1007605
Device Lot Number25424-1-1-1
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 05/18/2021
Initial Report FDA Received Date06/17/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Other;
Patient Age22 YR
-
-