| Model Number |
DL-330-10-K |
| Medical Device Problem Code |
Impedance Problem (2950)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/18/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).Neuropace evaluated the data uploaded to pdms (including ecogs and impedances).The review of the data was consistent with a potential lead break.The explanted product was discarded by the treating center and was not returned to neuropace for analysis.
|
| |
|
Event or Problem Description
|
|
The treating clinician performed a review of the patient's device data and identified abnormal impedances and artifact present on the recorded ecogs on (b)(6) 2020.Responsive stimulation was disabled on contact 2 of the affected lead.The patient's response was monitored until such time the clinician deemed a replacement of the affected lead was necessary.A lead replacement procedure was performed on (b)(6) 2021 without complication.There is no clear indication of head trauma or other inciting event.No other information was reported from the treating center.
|
| |
|