MAUDE Adverse Event Report: SMITH & NEPHEW, INC. ACCORD TENSIONER; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. ACCORD TENSIONER; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL Back to Search Results
Model Number 71360020
Medical Device Problem Code Failure to Reset (1532)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/01/2021
Type of Reportable Event Malfunction
Event or Problem Description
It was reported that an accord tensioner reset switch does not work consistently.As this was noticed in a non-surgical environment, there was not any patient involved.
 
Additional Manufacturer Narrative
H6: the device, intended for use in treatment, was returned and evaluated.A visual inspection of the returned accord tensioner confirms the device is bent in the middle of the handle.A functional evaluation of the returned product confirmed the stated failure mode.The device was paired with a mating part but could not function, rendering the device inoperable.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of complaint history on the listed part revealed no additional complaints for the listed batch.The device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.Should additional information be received, the complaint will be reopened.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCORD TENSIONER
Common Device Name
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12018352
Report Number1020279-2021-05219
Device Sequence Number3604486
Product Code JDG
UDI-Device Identifier03596010485359
UDI-Public03596010485359
Combination Product (Y/N)N
Initial Reporter StateKY
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71360020
Device Catalogue Number71360020
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/01/2021
Initial Date Received by Manufacturer 06/01/2021
Supplement Date Received by Manufacturer07/21/2021
Initial Report FDA Received Date06/17/2021
Supplement Report FDA Received Date07/21/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-