MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC TI PS PL MANDBL/ O ANG/2.0 THK; PLATE, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC TI PS PL MANDBL/ O ANG/2.0 THK; PLATE, BONE Back to Search Results
Model Number SD480.112
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Unspecified Infection (1930)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).Complainant device/part is not expected to be returned for manufacturer review/investigation, but has yet to be received.Not available - implanted.(b)(4).Investigation summary: product was not returned.Based on the information available, it has been determined that no corrective and preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device history lot: part number: sd480.112, lot number: 112p190, part manufacture date: 05-apr-2021, manufacturing location: (b)(4), part expiration date: n/a, nonconformance noted: n/a.Dhr record review: a review of the device history record revealed no complaint related anomalies.The device history record shows this lot of ti patient specific plate mandible/ w/o angle/2.0mm thk product was processed through the normal manufacturing and inspection operations with no rework nor nonconformities noted.This lot met all dimensional, visual and packaging criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.A review of the raw material device history record revealed this lot met all specifications with no nonconformance noted.This raw material lot met all dimensional and visual criteria at the time of manufacture with no issues documented during the manufacture that would contribute to this complaint condition.Device history batch null, device history review this lot met all dimensional, visual and packaging criteria at the time of release with no issues documented during the manufacture that would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that on an unknown date, the implant had to be removed due to infection.This complaint involves one (1) device.This report is for one (1) ti ps pl mandbl/ o ang/2.0 thk.This report is 1 of 2 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TI PS PL MANDBL/ O ANG/2.0 THK
Common Device Name
PLATE, BONE
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
MDR Report Key12018487
Report Number2939274-2021-03022
Device Sequence Number3688056
Product Code JEY
UDI-Device Identifier10887587005348
UDI-Public(01)10887587005348
Combination Product (Y/N)N
Initial Reporter StateTN
PMA/510(K) Number
K122647
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 05/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberSD480.112
Device Catalogue NumberSD480.112
Device Lot Number112P190
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/24/2021
Supplement Date Received by Manufacturer07/16/2021
Initial Report FDA Received Date06/17/2021
Supplement Report FDA Received Date07/16/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
Yes
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
UNK - SCREWS: CMF
-
-