• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MDT POWERED SURGICAL SOLUTIONS LEGEND ATTACHMENT; MOTOR, DRILL, PNEUMATIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MDT POWERED SURGICAL SOLUTIONS LEGEND ATTACHMENT; MOTOR, DRILL, PNEUMATIC Back to Search Results
Model Number AA07
Medical Device Problem Code Overheating of Device (1437)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Report confirmed.The device was tested and the max temperature was measured to be 160°f.The likely cause was identified as worn distal bearings.It was also noted that the color band was discolored and the laser markings were faded.The device user manual warnings section includes instructions that heavy side loads and/or long operating times may cause the device to overheat.If the device overheats, discontinue use and rest the device by using it intermittently, or wrap the device in a moist sterile towel.Use of an overheated device may cause injury to the patient or operator.We will continue to track and trend this complaint type.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Repair request initiated for device with the report of overheating.It was reported that there was no patient involvement.Repair request escalated to a product event based on reason for return and customer indication that this report is a complaint.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LEGEND ATTACHMENT
Common Device Name
MOTOR, DRILL, PNEUMATIC
Manufacturer (Section D)
MDT POWERED SURGICAL SOLUTIONS
4620 n beach st
fort worth TX 76137
MDR Report Key12021990
Report Number1625507-2021-00175
Device Sequence Number16487757
Product Code HBB
UDI-Device Identifier00643169740983
UDI-Public00643169740983
Combination Product (Y/N)N
Initial Reporter CountryIT
PMA/510(K) Number
K020069
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date (Section B) 06/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberAA07
Device Catalogue NumberAA07
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/30/2019
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 12/12/2019
Initial Report FDA Received Date06/17/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-