Model Number ZCU525
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Visual Impairment (2138)
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Weight and ethnicity: unknown, information not provided.Date of event: unknown, not provided, but the best estimate date is between (b)(6) 2021.Manufacturer telephone number: (b)(4).All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Event or Problem Description
It was reported that the intraocular lens (iol) was explanted from patient's right eye due to refractive surprise.Another lens of different model and different diopter was placed as the replacement.The reported event was observed during post-operative examination.No further information was available.
 
Additional Manufacturer Narrative
Additional information: section d9: device available for evaluation: yes.Section d9: returned to manufacturer on: july 19, 2021.Section h3: device evaluated by manufacturer: yes.Device evaluation: the product was returned to the manufacturing site for evaluation.Visual inspection under magnification revealed viscoelastic residue on the optic body and haptics.The lens was cleaned, no cosmetic issues were observed on the optic body.The complaint issue could not be confirmed, and no product deficiency could be identified.Manufacturing record review: the manufacturing records for the intraocular lens were reviewed.The product was manufactured and released according to specification.The search revealed no additional investigation request (ir) for this production order number has been received.Conclusion: as a result of the investigation there is no indication of a product quality deficiency.All pertinent information available to johnson & johnson surgical vision, inc.Has been submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECNIS IOL
Common Device Name
INTRAOCULAR LENS
Manufacturer (Section D)
AMO MANUFACTURING NETHERLANDS
van swietenlaan 5
groningen, groningen 9728 NX
NL  9728 NX
MDR Report Key12022650
Report Number9614546-2021-07177
Device Sequence Number14537874
Product Code HQL
UDI-Device Identifier05050474664098
UDI-Public(01)05050474664098(17)250806
Combination Product (Y/N)N
Initial Reporter StateWA
Initial Reporter CountryUS
PMA/510(K) Number
P980040
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source health professional,user faci
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 08/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date08/06/2025
Device Model NumberZCU525
Device Catalogue NumberZCU525U105
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/19/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 05/21/2021
Supplement Date Received by Manufacturer08/10/2021
Initial Report FDA Received Date06/18/2021
Supplement Report FDA Received Date08/20/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age70 YR
-
-