As the lot number for the device was provided, a review of the device history record is currently being performed.The device has been returned to the manufacturer for evaluation.However, a photo was provided for review.The investigation of the reported event is currently underway.(expiry date: 11/2021).
|
|
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one 14s crosser cto recanalization catheter was returned for evaluation.Device appeared to have residue throughout.No kinks noted to the catheter, no anomalies noted to transducer handle, or saline hub.Marker band is present and undamaged.Material separation also noted just proximal to the distal tip.Distal tip was present and attached.Based on the findings, the reported material separation issue can be confirmed as a tip separation noted to the device.Additionally, one electronic photo was provided and reviewed.The photo shows the distal tip of the crosser catheter.The tip separation from the catheter can be noted to the device.Therefore, based on the photo review, the reported material separation issue can be confirmed as a tip separation noted to the device.A definitive root cause for the reported material separation issue could not be determined based upon the provided information.Labeling review: a review of product labeling documentation (e.G., procedural instructions, indications, warnings, precautions, cautions, possible complications, contraindications, nursing guide, and unit label) did not find any product labeling inadequacy.H10: d4 (expiry date: 11/2021), g3, h6 (method).H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
|