| Model Number |
7580 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/23/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Age at time of event - above 18 years and older.Device evaluated by mfr.: returned product consisted of a maverick 2 mr balloon catheter.The device was microscopically and visually examined.At 32.8cm from the strain relief the hypotube of the device was kinked.There was a minimal amount of blood present in the guidewire lumen.The balloon was loosely folded.Spanning 14mm long into both proximal and distal balloon cone-waist transitions there was a longitudinal tear.The device presented no other damages or irregularities.
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on (b)(6) 2021.It was reported that balloon damage occurred.A 2.00mm x 12mm maverick balloon catheter was selected to treat a patient with coronary artery disease.It was noted that the balloon was damaged.The device was replaced with another of the same device.No patient complications were reported and the patient status was stable.However, returned device analysis revealed balloon longitudinal tear.
|
| |
|