Model Number 1363-47-000
Medical Device Problem Code Device Contaminated During Manufacture or Shipping (2969)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/04/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
During opening, the material sticker was already on the implant - the customer did not want do implant that.A new implant was opened.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: b3.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Additional Manufacturer Narrative
Product complaint #
=
> (b)(4) investigation summary
=
> examination of the returned device did not identify any product defects or anomalies.Depuy considers the investigation closed.Should additional info be received, the investigation will be re-opened as necessary.Device history lot
=
> a search of the depuy nonconformance (nc) quality system found no nc¿s associated with this product/lot combination.Corrected: h3.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MODULAR CATHCART BALL 47MM OD
Common Device Name
HEMI HIP IMPLANT : HIP ACETABULAR METAL HEMI
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key12025496
Report Number1818910-2021-12995
Device Sequence Number17228201
Product Code LZY
UDI-Device Identifier10603295032793
UDI-Public10603295032793
Combination Product (Y/N)N
Initial Reporter CountryAU
PMA/510(K) Number
K903084
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup
Report Date (Section B) 06/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number1363-47-000
Device Catalogue Number136347000
Device Lot NumberD20102089
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/19/2021
Initial Date Received by Manufacturer 06/04/2021
Supplement Date Received by Manufacturer06/14/2021
07/19/2021
08/04/2021
Initial Report FDA Received Date06/18/2021
Supplement Report FDA Received Date06/22/2021
07/22/2021
08/04/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-