Model Number 9733856 |
Device Problem
Excessive Heating (4030)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/28/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
No patient information provided as no patient was involved when the reported issue occurred.A medtronic representative went to the site to test the equipment.Testing revealed that the reported issue could not be replicated.However, while servicing there were display issues noted on the system as the monitor booted to have no signal and also powered off multiple times without prompt from the user.Subsequently the monitor cable was replaced.The navigation system then passed the system checkout and was found to be fully functional.Concomitant medical products: other relevant device(s) are: product id: 9 733571, serial/lot #: (b)(4), ubd:, udi#:.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information regarding a navigation device being used outside of a procedure.It was reported that the surgeon monitor cable was running hot.It was noted that the cable did not appear to be damaged.There was no patient present when this issue was identified.
|
|
Manufacturer Narrative
|
H3) the monitor cable for the navigation system was returned to the manufacturer for evaluation.Testing found that the cable had bent and twisted pins in a connector.Continuity testing found an open to ground on one pin.The hardware investigation found that the reported event was related to a hardware issue.This issue was documented in a medtronic navigation hardware anomaly tracking database.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|