• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WRIGHTS LANE SYNTHES USA PRODUCTS LLC 2.5MM DRILL BIT/QC/GOLD/110MM; INSTR,SURGICAL,ORTHOPEDIC,AC- POWERED MOTOR/ACCESS & ATTACH

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHTS LANE SYNTHES USA PRODUCTS LLC 2.5MM DRILL BIT/QC/GOLD/110MM; INSTR,SURGICAL,ORTHOPEDIC,AC- POWERED MOTOR/ACCESS & ATTACH Back to Search Results
Model Number 310.25
Device Problems Break (1069); Entrapment of Device (1212)
Patient Problem Foreign Body In Patient (2687)
Event Date 05/24/2021
Event Type  Injury  
Manufacturer Narrative
Product complaint # (b)(4).Additional product code: gff gfa hsz.Complainant part is not expected to be returned for manufacturer review/ investigation.Reporter is a j&j sales representative.Investigation summary; photo investigation: the device was not returned.A photo-investigation was performed on the images.Upon inspecting the images provided, the 2.5mm drill bit was observed to be broken at the tip.However, the allegation of the broken part of the drill bit being embedded in the patient cannot be confirmed through the provided x-ray images.The root cause for the reported issue cannot be determined from the available information.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.Conclusion: the complaint condition can be confirmed during photo investigation.During the investigation, no product design issues or discrepancies were observed.No manufacturing issues were noted during investigation.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post market safety surveillance activities.Therefore, it has been determined that no corrective and/or preventive action is proposed.Device history lot part #: 310.25.Synthes lot #: u368972.Supplier lot #: u368972.Release to warehouse date: oct, 21 2020.Supplier: orchid unique.No ncr's generated during production.Device history review: review of the device history record(s) showed that there were no issues during the manufacture of this product, and any sub-components, which would contribute to this complaint condition.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that on (b)(6) 2021, the patient underwent for a surgery.During the surgery, a 2.5 drill bit broke and the tip was left in patient distal femur.It is unknown if there was a surgical delay.Procedure and patient outcome were unknown.This complaint involves (1) device.This report is for (1) 2.5mm drill bit/qc/gold/110mm.This report is 1 of 1 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2.5MM DRILL BIT/QC/GOLD/110MM
Type of Device
INSTR,SURGICAL,ORTHOPEDIC,AC- POWERED MOTOR/ACCESS & ATTACH
Manufacturer (Section D)
WRIGHTS LANE SYNTHES USA PRODUCTS LLC
1302 wrights lane east
west chester PA 19380
Manufacturer (Section G)
MONUMENT
1101 synthes avenue
monument CO 80132
Manufacturer Contact
kara ditty-bovard
1302 wright lane east
west chester, PA 19380
6107195000
MDR Report Key12026129
MDR Text Key263859119
Report Number2939274-2021-03049
Device Sequence Number1
Product Code HWE
UDI-Device Identifier10886982186317
UDI-Public(01)10886982186317
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/25/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/18/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number310.25
Device Catalogue Number310.25
Device Lot NumberU368972
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/25/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/21/2020
Is the Device Single Use? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age73 YR
Patient Weight103
-
-