| Model Number |
212518 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/24/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Date of event: unknown.The date received by manufacturer has been used for this field.Investigation summary: components are mixed and dispensed into the appropriate containers.After qc testing product is released and transported to the distribution center.The complaint investigation included a review of the batch history record for tb carbolfuchsin kf.The batch history record reviews indicated no discrepancies.All release testing was satisfactory.Complaint trends were reviewed for a period covering 12 months.During that time there have been no confirmed complaints for the defect in question.One photo was received.The photo depicts a microscopic view of a stain slide.A mixture of purple and blue green stain can be seen.A retention sample from the complaint batch was evaluated along with a control batch.The solution appearance was satisfactory and comparable to the control; both were a reddish-purple suspension with no visible precipitate, impurities, or contamination.Staining was completed per instructions in the product insert.Slides of positive (m.Tuberculosis atcc 25177) and negative (b.Subtilis atcc 6533) controls were evaluated and gave satisfactory staining results with the retention and control batches.The retention was comparable to the control.This complaint is not confirmed.
|
| |
|
Event or Problem Description
|
|
It was reported that while using bd bbl¿ tb carbolfuchsin kf contamination was observed by the laboratory personnel.There was no indication that results were reported out and there was no report of patient impact.The following information was provided by the initial reporter: "impurities".
|
| |
|
Search Alerts/Recalls
|