| Model Number |
45-480820-L |
| Medical Device Problem Code |
Detachment of Device or Device Component (2907)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/19/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
A search for non-conformances associated with the reported part/lot number combination did not reveal any production-related issues relevant to the complaint that occurred during manufacturing of the device.The device was discarded at the user facility and not returned to the manufacturer for analysis.The alleged product issue could not be confirmed.The instructions for use (ifu) identifies premature coil detachment as a potential complication associated with use of the device.
|
| |
|
Event or Problem Description
|
|
It was reported that an embolization coil implant detached in the microcatheter.The implant was removed in its entirety along with the microcatheter and it was flushed out on the table.There was no reported intervention or patient injury.
|
| |
|