| Model Number |
NICE1 |
| Medical Device Problem Code |
Patient Device Interaction Problem (4001)
|
| Health Effect - Clinical Codes |
Necrosis (1971); Tissue Breakdown (2681); Swelling/ Edema (4577)
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
The device was not returned for investigation.However, the device was evaluated by a third-party group.No faults were found and it appears to be functioning properly.The dhr was reviewed and no anomalies or discrepancies were identified that would contribute to the reported event.Should the device be returned for investigation, additional testing will be performed on the wrap a follow-up report will be submitted at that time.No additional containment steps have been implemented at this time.
|
| |
|
Event or Problem Description
|
|
It was reported that a patient was hospitalized for 4 weeks and diagnosed with compartment syndrome induced by cryotherapy after using the device status post bunionectomy and first metatarsal osteotomy.The wrap was built into the surgical dressing after putting layers of web roll on the skin and compression was not used.The patient underwent hyperbaric oxygen treatment while he was hospitalized but no further information is available at this time.
|
| |
|