| Model Number |
AQL-100CS |
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
05/14/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The most probable root cause was either (1) a damaged or a de-calibrated weighing cell, (2) objects placed on the containers or (3) fluid on the under-buttocks-drape remaining and afterwards the deficit was zeroed out on the device.Affected devices: pump serial (b)(4).Cart serial (b)(4).
|
| |
|
Event or Problem Description
|
|
Technical support was contacted by (b)(6), territory manager, to report customer owned aquilex system is showing a negative deficit and is due for a pm.
|
| |
|