• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MPRI ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MPRI ACTIVA; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS Back to Search Results
Model Number 3387S-40
Medical Device Problem Codes Low impedance (2285); Malposition of Device (2616); Difficult to Open or Close (2921); Human-Device Interface Problem (2949)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 06/04/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that the physician did a bilateral subthalamic nucleus (stn) lead placement and expressed that he had difficulty making the clip snap into the stimloc base.They did not visibly see any excess nylon molding.There were no factors that may have led or contributed to the issue.New stimlocs were offered but the physician declined and stated that he was able to force the caps into the stimloc base.Additional information was received from a healthcare provider (hcp), via the manufacturer¿s representative (rep) who reported during a stage 2 case the lead was checked using the twistlock cable and it was found there were low impedances on contact pairings 3/0 and 2/1 (ranging from 91-104).The surgeon removed the lead from the twistlock cable, inspected it, and wiped it down with a wet then dry sponge.The surgeon tested again with the same results.Impedances were tested on 0.7, 1.5, and 3.0 v with the same results with the lead appearing to be intact.The surgeon the connected the extension wire and neurostimulator,and the same results appeared during the impedance check.The surgeon decided to the leave the lead as is and considered replacing it in the future.It was noted that during the stage 1 case the surgeon pushed the lead down deeper due to it being found to be 3 mm superior to the intended plan an d it was suspected this could have potentially unknowingly caused the impedance issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACTIVA
Common Device Name
STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
MDR Report Key12033432
Report Number2649622-2021-12479
Device Sequence Number20025017
Product Code MHY
UDI-Device Identifier00643169752290
UDI-Public00643169752290
Combination Product (Y/N)N
Initial Reporter StateNV
Initial Reporter CountryUS
PMA/510(K) Number
P960009
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 06/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date03/09/2024
Device Model Number3387S-40
Device Catalogue Number3387S-40
Device Lot NumberVA2E93K
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/17/2021
Initial Report FDA Received Date06/21/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age64 YR
-
-