| Model Number |
PED-350-12 |
| Medical Device Problem Codes |
Break (1069); Unintended Ejection (1234)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
06/16/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Medtronic received information regarding a pipeline pushwire that was thought to have separated.The patient was undergoing a procedure for flow diverter treatment of an unruptured saccular aneurysm in the left internal carotid artery (ica).The aneurysm max diameter was 3mm and the neck diameter was 2mm.Vessel tortuosity was minimal.It was reported that the pipeline and all accessory devices were prepared as indicated in the instructions for use (ifu).During deployment of the pipeline, the physician felt that the pushwire detached from the device so it was decided to recapture the pipeline and start over.It was noted there was no friction or other difficulty but thought the pushwire may have been damaged.Another pipeline was implanted successfully.There was no harm or injury to the patient.Ancillary device: phenom 27 microcatheter.
|
| |
|
Event or Problem Description
|
|
Additional information received indicated there was no pushwire break or apparent damage.The device seemed to disconnect from the pushwire.
|
| |
|
Additional Manufacturer Narrative
|
|
H10: review of this mdr and/or additional information received shows that there is no information to reasonably suggest that the device in this report may have caused or contributed to a death or serious injury or that the device in this report has malfunctioned in a way that may have caused or contributed to a death or serious injury.Therefore, this event did not and does not meet the reporting requirements stipulated in 21 cfr 803." medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Search Alerts/Recalls
|