• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ACE SURGICAL SUPPLY COMPANY INTERNAL HEX IMPLANT; DENTAL IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ACE SURGICAL SUPPLY COMPANY INTERNAL HEX IMPLANT; DENTAL IMPLANT Back to Search Results
Model Number 204711
Medical Device Problem Codes Failure to Osseointegrate (1863); Osseointegration Problem (3003)
Health Effect - Clinical Code Failure of Implant (1924)
Date of Event 05/20/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
The reported event has been determined to be a post-placement/pre-load failure.This implant loss suggests that the source of the problem was likely to stem from no osseointegration.Additionally, other factors that may have contributed to the implant failure includes bone condition, patient oral hygiene, overall health or behavior.Proper intended use of the implant is described in its instructions for use.
 
Event or Problem Description
Wrong size implanted was reported.Bone quality at the time of implant failure type iii.Time of loss in relation to the implantation was before prosthetic restoration.Primary stability and osseointegration was not achieved.Augmentation procedure was not performed at the time of surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
INTERNAL HEX IMPLANT
Common Device Name
DENTAL IMPLANT
Manufacturer (Section D)
ACE SURGICAL SUPPLY COMPANY
1034 pearl street
brockton MA 02301
MDR Report Key12035178
Report Number1287163-2021-00887
Device Sequence Number11862961
Product Code DZE
UDI-Device Identifier00614950004751
UDI-Public00614950004751
Combination Product (Y/N)N
Initial Reporter StateNJ
Initial Reporter CountryUS
PMA/510(K) Number
K102981
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Device Explanted Year2021
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Dentist
Type of Report Initial
Report Date (Section B) 06/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Model Number204711
Device Catalogue Number204711
Device Lot Number20090002
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/15/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 06/15/2021
Initial Report FDA Received Date06/21/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age68 YR
-
-