Model Number 260314
Medical Device Problem Code Product Quality Problem (1506)
Health Effect - Clinical Code Needle Stick/Puncture (2462)
Date of Event 05/28/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product involved in incident was not returned for evaluation.If product is returned, arkray will send to manufacturer for testing and file a follow-up report when evaluation is complete.
 
Event or Problem Description
Part 260314.Lot 02909043.Exp date: 2024-10-20.Caller is claiming that she has four boxes of insulin syringes and that many of the syringes are uncapped within the packaging that holds the syringe and that many of the packages that hold the syringes are of low quality in her opinion and are ripped or damaged.Caller also claimed that she was showing a customer how to use the insulin syringe and that she was accidentally stuck by the needle during the demonstration and that the needle that stuck her was bent.Upon visual inspection of other needles in the packages she did not know if anymore were bent but claimed they could be bending because the caps are off on a lot of them.Since the caller was stuck with the needle, she did seek medical treatment as that is her company's policy for such an incident, and she did have a blood test performed to rule out any possible infections through blood borne pathogens, but the needle she was stuck with was a brand new needle out of the box and she was the only person who had used it, so the needle was sterile when it poked her.Caller stated that she did not sell any of these products to the customer that she was demonstrating it to, because she became concerned about the quality of the product.When offered a replacement for the boxes of syringes in question, caller refused replacement as she has decided that she will not be offering this brand of syringes to her customers.Caller will return the complaint boxes of syringes for further investigation.
 
Additional Manufacturer Narrative
Complaint devices involved in complaint were returned on 6/29/2021 and forwarded to the manufacturer for further investigation.Investigation results were received on 2/8/2022.Retains of devices were inspected and no anomalies were found.10 samples were returned by customer.7 out of 10 were out of product specification requirement with the cap not on the needle.Root cause was conducted, and the issue was determined to be caused by long-distance transportation vibration.Manufacturer has implemented a 100% check of product for cap position after package is sealed.This will help confirm all the products were capped and assembled properly before shipment.Risk assessment analysis is low based on sales of product and overall complaints, of which in the last year, there were three complaints related to this issue.No trend evidence.Complaints will continue to be monitored for indication of possible trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TECHLITE INSULIN SYRINGE
Common Device Name
SYRINGE, PISTON
Manufacturer (Section D)
SOL-MILLENNIUM MEDICAL INC.
315 shawnee drive n suite 100
suwanee GA 30024
Manufacturer (Section G)
SOL-MILLENNIUM MEDICAL INC.
315 shawnee drive n suite 100
suwanee GA 30024
Manufacturer Contact
tom speikers
5182 w 76th street
minneapolis, MN 55439
9526463168
MDR Report Key12035517
Report Number1832816-2021-00013
Device Sequence Number4166585
Product Code FMF
UDI-Device Identifier00015482613143
UDI-Public(01)00015482613143
Combination Product (Y/N)N
Initial Reporter StateVA
Initial Reporter CountryUS
PMA/510(K) Number
K153537
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source Health Professional
Initial Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date (Section B) 02/14/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date10/20/2024
Device Model Number260314
Device Catalogue Number260314
Device Lot Number02909043
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/29/2021
Is the Reporter a Health Professional? Yes
Type of Report(Section G)Follow-Up
Initial Date Received by Manufacturer 05/29/2021
Supplement Date Received by Manufacturer05/29/2021
Initial Report FDA Received Date06/21/2021
Supplement Report FDA Received Date02/14/2022
Was Device Evaluated by Manufacturer? (Y/N) Yes
Date Device Manufactured12/02/2019
Is the Device Labeled for Single Use? (Y/N) Yes
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Usage of Device Initial
Patient Sequence Number1
Patient Age37 YR
Patient SexFemale
Patient Weight50 KG
-
-